The Model 1057 analyzer offers three sensor inputs and four current outputs thus reducing the cost per loop
and saving panel space. The pH signal input board supports pH, ORP, and Ion-Selective Electrode measurements.
The Conductivity signal input board supports contacting conductivity, resistivity, Total Dissolved
Solids, salinity and percent concentration curves for special applications. The modular design allows signal
input boards to be field replaced making configuration changes easy. Conveniently, live process values are
always displayed during programming and calibration routines. Standard features include isolated inputs, 8
embedded local languages, four 4-20mA current outputs, 4 alarm relays and removable connectors
for power and current outputs. QUICK START PROGRAMMING: Exclusive Quick
Start screens appear the first time the Model 1057 is powered. The instrument auto-recognizes each
measurement board and prompts the user to configure each sensor loop in a few quick steps for immediate deployment.
SPECIFICATIONS - General
Enclosure: Polycarbonate. NEMA 4X/CSA 4 (IP65).
Dimensions: Overall 155 x 155 x 131mm (6.10 x 6.10 x
5.15 in.). Cutout: 1/2 DIN 139mm x 139mm (5.45 x 5.45 in.)
Minimum depth for panel mount installations 101.6 mm
(4.0 in).
Conduit Openings: Accepts 1/2” or PG13.5 conduit
fittings
Display: Monochromatic graphic liquid crystal display.
128 x 96 pixel display resolution. Backlit. Active
display area: 58 x 78mm (2.3 x 3.0 in.).
Ambient Temperature and Humidity: 0 to 55°C
(32 to 131°F). RH 5 to 95% (non-condensing)
Storage Temperature Effect: -20 to 60ºC (-4 to 140°F)
Power: Code -02: 20 to 30 VDC. 15 W.
Code -03: 84 to 265 VAC, 47.5 to 65.0 Hz, switching.
15 W
Model 1057 can also derive an inferred pH value called
pHCalc (calculated pH). pHCalc can be derived and
displayed when two contacting conductivity sensors are
used as sensor 1 and sensor 2.
PERFORMANCE SPECIFICATIONS -
ANALYZER (pH INPUT)
Measurement Range [pH]: 0 to 14 pH
Accuracy: ±0.01 pH
Diagnostics: Glass impedance, reference impedance
Temperature coefficient: ±0.002pH/ ºC
Solution temperature correction: Pure water, dilute
base and custom.
Buffer recognition: NIST, DIN 19266, JIS 8802, and BSI.
Input filter: Time constant 1 - 999 seconds, default 4
seconds.
Response time: 5 seconds to 100%
Sensor Compatibility: Model 1057 is also compatible
with SMART pH sensors from Rosemount
Analytical. Choose from 14 SMART sensor models
for a wide array of applications and process conditions.