he new Clarity II™ turbidimeter Model T1056 is intended for the determination of turbidity in water. Low stray light, high stability, efficient bubble rejection, and a display resolution of 0.001 NTU make Clarity II™ ideal for monitoring the turbidity of filtered drinking water. 

The Clarity II™ Model T1056 is a complete system. It includes an analyzer (single or dual input), sensor(s), debubbler/measuring chamber(s), and a cable for each sensor. The system complies with ISO method 7027 when used with the ISO/LED sensor and with EPA method 180.1 when used with the EPA/incandescent sensor. Clarity II uses a two-stage debubbler that reduces bubble accumulation in the measured water sample to prevent erroneous turbidity readings. 

The Clarity II™ turbidimeter incorporates the popular and easy to use Model 1056 analyzer. The Model 1056 analyzer features a large display with easy to read process variables, a modular design with slide-in measurement boards, and extensive diagnostics and User Help screens to assist in troubleshooting. Menu flows and prompts are so intuitive that a manual is practically not needed.
SPECIFICATIONS — DEBUBBLER AND
FLOW CHAMBER
Dimensions: 18.1 in. x 4.1 in. diam. (460 mm x 104 mm
diam.) (approx.)
Wetted materials: ABS, EPDM, Delrin1, polypropylene,
nylon
Inlet: compression fitting accepts 1/4 in. OD tubing; fitting
can be removed to provide 1/4 in. FNPT
Drain: barbed fitting accepts 3/8 in. ID tubing; fitting can
be removed to provide 1/4 in. FNPT. Must drain to
atmosphere.
Sample temperature: 40 95°
F (5 35°
C)
Minimum inlet pressure : 3.5 psig (125 kPa abs). 3.5
psig will provide about 250 mL/min sample flow.
Maximum inlet pressure: 30 psig (308 kPa abs). Do not
block drain tube.
Recommended sample flow: 250 750
mL/min
Response Time: The table shows the time in minutes to
percent of final value following a step change in turbidity.